The FMX Generator intelligently delivers electrical energy to the surgical instruments. The advanced generator features a patented software control algorithm that continuously monitors and adjusts the delivery of energy, ensuring that the optimal amount of heat is delivered to the tissue at all times. The large touchscreen display provides the entire surgical team immediate feedback on power setting levels and changing settings is as easy as touching the display. The FMX Generator has earned the Type CF (“Cardiac Floating”) rating.
Laser-Like Precision
- Tactile control with minimal tissue drag
- Surface only thermal effect
- Predictable, char-free layer-by-layer dissection with optimal visualization of tissue planes
- Safe to use near nerves, vessels, and all delicate organs, including the heart
Minimal Thermal Injury
- As little as 1/10th the thermal injury of monopolar electrosurgery
- As few as 80 microns (0.08 mm) of thermal spread
- Clear margins for reliable pathology specimens
- Surgeons note less unintended damage to tissue, leading to reduced use of blood products during surgery, and less post-operative edema and drainage
Electrical Silence
- No electrical current passes through tissue
- No grounding pad
- No spark, arcing, or stray current
Prepare the FMwand by following these simple steps:
- Plug the FMX Power Module into the front panel of the FMX Generator
- Snap the power module into the back of the FMwand Handpiece
- Set the power settings on the generator and the FMwand is ready
FMX Generator Technical Specifications
- Operating Environment Conditions
Temp: 10° to 30° C (50° to 86° F)
Humidity: 15% to 85% Non-condensing
- Storage Conditions
Temp: -20° to 65° C (-4° to 149° F)
Humidity: 15% to 85% Non-condensing
- Shipping Conditions
Temp: -20° to 65° C (-4° to 149° F)
Humidity: 15% to 85% Non-condensing
- Moisture Protection
Generator: IPX1
Foot Pedal: IPX7 minimum
- Generator Dimensions : 14″ x 4″ x 12″ 35.6 cm x 10.2 cm x 30.5 cm
- Generator Weight : Approximately 20 lbs. (9 kg)
- Power Requirements
Voltage: 110/220/230 Vac
Current: 4/2A rms max
Frequency: 50/60 Hz
- Power Output : 60W Max
- Fuses
For 110V: T 5A L 250V
For 220V/230V: T 2.5A L 250V
- Operating Frequency : 40.68 MHz
- Mode of Operation : Intermittent Operation 10 sec. ON, 30 sec. OFF
- Protection Class : Class I
- Output Type : Type CF